Model 425 VUV Channel Electron Multiplier
  • Windowless, useful for wavelengths ~1 to 180-nm
  • Solar blind
  • CsI, MgF2 and uncoated photocathode
  • 5" 'McPherson' flange mount
Model 425 Channel Electron Multiplier

The Model 425 Channel Electron Multiplier is a bakeable, windowless,solar blind detector for operation in the 1 to 180-nm region. It is mounted to a special ceramic board and encased in a vacuum tight housing for vacuum operation. The Model 425 is ideal for measurements in the Extreme and Vacuum UV (EUV and VUV) where the solar blind feature eliminates background noise. It is especially suitable for photon counting applications. The CEM is also available with a variety of coatings (or uncoated) to enhance response in different energy regions.

The vacuum housing fits directly to the exit slit of your McPherson monochromator or alternately, can be installed directly inside some McPherson vacuum sample chambers. The detector assembly includes two auxiliary pipe thread ports (1/8" NPT) for use with vacuum gauges and/or supplemental pumping. Vacuum feed throughs for operation to 3,000 VDC and for signal output are provided.

The Model 7642 is a suitable 5-kV high voltage power supply for proper operation of this detector.

The CEM will be permanently damaged if operated at vacuum pressures exceeding 1 x 10^4 torr. Interlocking the vacuum gage system to the detectors HV supply is strongly recommended

Part Numbers for Model 425 Channel Electron Multiplier Assemblies
Part NumberDescription
8105-0090-1 Model 425 channel electron multiplier with vacuum housing and Cesium Iodide (CsI) coated detector for best response in the 1 to 180-nm region. Provided with a mounting flange fitting McPherson slits.
8105-0090-2 Model 425 channel electron multiplier with vacuum housing and Magnesium Fluoride (MgF2) coated detector for best response in the 1 to 120-nm region. Provided with a mounting flange fitting McPherson slits.


Model 425 photon sensitivity with CsI and Uncoated channel electron multiplier

Model 425 photon sensitivity with CsI and Uncoated channel electron multiplier


...Back to Detector Index...



Sales@McPhersonInc.com, Telephone (978)256-4512, Fax (978) 250-8625
McPherson, Inc. - 7A Stuart Road, Chelmsford, MA 01824 USA